Health Products

VITROS Immunodiagnostic Products AFP Reagent Pack and Calibrators (2016-05-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, General Public, Healthcare Professionals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
1440A. VITROS Immunodiagnostic Products AFP Reagent Pack and Calibrators
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
FELINDRE MEADOWS, PENCOED, BRIDGEND
1515154

Introduction

Affected products

A. VITROS Immunodiagnostic Products AFP Reagent Pack and Calibrators

Reason

Ortho Clinical Diagnostics (Ortho) has initiated this recall due to the potential of an inability to obtain a valid calibration when using VITROS AFP Reagent Packs:
- VITROS Immunodiagnostic Products AFP Reagent Pack, lot 1440
- VITROS Immunodiagnostic Products AFP Calibrators, lot 1440
Ortho confirmed that this lot exhibited an increase in reaction signal over time. Initial testing indicates that the unexpected signal increase is related to a specific lot of raw material used in the manufacture of this lot. The affected lot of VITROS AFP Reagent Packs met all quality assurance specifications upon release.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime