A. VITROS Immunodiagnostic Products AFP Reagent Pack and Calibrators
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
FELINDRE MEADOWS, PENCOED, BRIDGEND
1515154
Introduction
Affected products
A. VITROS Immunodiagnostic Products AFP Reagent Pack and Calibrators
Reason
Ortho Clinical Diagnostics (Ortho) has initiated this recall due to the potential of an inability to obtain a valid calibration when using VITROS AFP Reagent Packs:
- VITROS Immunodiagnostic Products AFP Reagent Pack, lot 1440
- VITROS Immunodiagnostic Products AFP Calibrators, lot 1440
Ortho confirmed that this lot exhibited an increase in reaction signal over time. Initial testing indicates that the unexpected signal increase is related to a specific lot of raw material used in the manufacture of this lot. The affected lot of VITROS AFP Reagent Packs met all quality assurance specifications upon release.