D. DIMENSION EXL SYSTEM-ENZYME II CALIBRATOR (ENZ II CAL)
10476170
4MD047
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Alabama
UNITED STATES
B. DIMENSION CLINICAL CHEMISTRY SYSTEM- ENZYME II CALBIBRATOR (ENZ II CAL)
10476170
4MD047
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Alabama
UNITED STATES
C. DIMENSION XPAND SYSTEM-ENZYME II CALIBRATOR (ENZ II CAL)
10476170
4MD048
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Alabama
UNITED STATES
A. DIMENSION VISTA SYSTEM- ENZYME CALIBRATOR
10475527
Introduction
Affected products
A. DIMENSION VISTA SYSTEM- ENZYME CALIBRATOR
B. DIMENSION CLINICAL CHEMISTRY SYSTEM- ENZYME II CALBIBRATOR (ENZ II CAL)
C. DIMENSION XPAND SYSTEM-ENZYME II CALIBRATOR (ENZ II CAL)
D. DIMENSION EXL SYSTEM-ENZYME II CALIBRATOR (ENZ II CAL)
Reason
Siemens Healthcare Diagnostics has received complaints of positive shift of patient and QC results when the ALTI assay is calibrated with specific lots of Dimension Enzyme II Calibrator. Internal investigation has confirmed that patients, quality control, and proficiency testing material all demonstrate a similar shift of up to 10% in comparison to results obtained using earlier calibration lots. The shift began with lot 4MDO47.