Health Products

ALERE TESTPACK+PLUS STREP A WITH OBC (2016-05-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
55331K100
Manufacturer
ALERE MEDICAL CO., LTD.
357 MATSUHIDAI
MATSUDO, CHIBA-KEN
270-2214
JAPAN
A. ALERE TESTPACK+PLUS STREP A WITH OBC505715J

Introduction

Affected products

A. ALERE TESTPACK+PLUS STREP A WITH OBC

Reason

Affected devices contain swabs that are defective. The defective swab shafts are yellow in colour when normally white and can be easily broken if bent. In cases where the swab is broken during sample collection from the throat, a patient's throat may be injured and/or the swab may be swallowed.