ABX PENTRA 400 CLINICAL CHEMISTRY SYSTEM (2016-02-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
361014384
Manufacturer
Horiba ABX, Parc Euromedecine
rue Du Caducee - BP 7290, Cedex 4,
Montpellier
34184
FRANCE
A. ABX PENTRA 400 CLINICAL CHEMISTRY SYSTEMP4000000NUA1

Introduction

Affected products

A. ABX PENTRA 400 CLINICAL CHEMISTRY SYSTEM

Reason

A malfunction with reagent dispensing on the ABX Pentra 400 (version 5.0.8 & lower) and Pentra C400 (version 1.1.2 & lower) chemistry analyzers for the applications using 2 or more reagents in their assay reactions. Could potentially lead to falsely low or falsely high results.