Health Products

PHYSIOMESH FLEXIBLE COMPOSITE MESH (2016-05-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
JOHNSON & JOHNSON INTERNATIONAL, C/O EUROPEAN LOGISTICS CENTRE
LEONARDO DA VINCILAAN, 15, DIEGEM
BELGIUM
A. PHYSIOMESH FLEXIBLE COMPOSITE MESHPHY0715R

Introduction

Affected products

A. PHYSIOMESH FLEXIBLE COMPOSITE MESH

Reason

Ethicon is recalling the product following an analysis of unpublished data from two large independent hernia registries. Recurrence / reoperation rates (respectively) after Laparoscopic Ventral Hernia Repair using ETHICON PHYSIOMESH Flexible Composite Mesh were higher than the average rates of the comparator set of meshes among patients in these registries. Ethicon is unable at this time to characterize all factors contributing to higher rates and is therefore removing the product from the global market.