Health Products

ALDOLASE MANUAL METHOD (2016-05-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Randox Laboratories Ltd.
55 Diamond Road
Ardmore, County Antrim
BT29 4QY
UNITED KINGDOM
A. ALDOLASE MANUAL METHODAD189

Introduction

Affected products

A. ALDOLASE MANUAL METHOD

Reason

Customers using a Randox RX imola instrument only, will not be able use Randox Aldolase reagent (AD189) with the instrument specific application provided. The parameters shown on the screen shots on the instrument specific application sheet for Aldolase assay on the RX imola instrument are incorrect. The settings do not allow for absorbance change, therefore calibration is not possible, and the assay cannot be run as a test. Affected customers will receive an updated instrument specific application sheet for Aldolase with the correct parameters.