Health Products

BrightView Imaging Systems (2016-05-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES
B. SYST, BRIGHTVIEW X WITH CRYSTAL453560829241
Not applicable
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES
A. BRIGHTVIEW XCT IMAGING SYSTEM453560462131

Introduction

Affected products
  1. BRIGHTVIEW XCT IMAGING SYSTEM
  2. SYST, BRIGHTVIEW X WITH CRYSTAL

Reason

During a Non-auto Body Contouring (abc), non-circular, REL-180 scan, the detector may come in contact with the patient. If this occurs, contact sensors would pause the scan.