VISUALASE COOLED CATHETER LASER APPLICATOR (2016-05-27) May 27, 2016 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number All lots Manufacturer Medtronic Navigation Inc. 826 Coal Creek Circle Colorado UNITED STATES A. VISUALASE COOLED CATHETER LASER APPLICATOR T10: 001-4000 - VCLAS 001-4000 .4MM CORE FIBER 10M
Affected products
A. VISUALASE COOLED CATHETER LASER APPLICATOR
Manufacturer received reports of melting of the VCLAS catheters during procedures to ablate tissue. Manufacturer revised the recommended treatment parameters and revised and added warnings accordingly.