Health Products

VITROS Immunodiagnostic Products Total T4 Reagent Pack (2016-06-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, Healthcare Professionals, General Public
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
1836A. VITROS Immunodiagnostic Products Total T4 Reagent Pack
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
FELINDRE MEADOWS, PENCOED, BRIDGEND
8744468

Introduction

Affected products

A. VITROS Immunodiagnostic Products Total T4 Reagent Pack

Reason

Ortho Clinical Diagnostics (Ortho) has initiated this recall because it was determined that the open pack (in use) storage limits must be reduced from 8 weeks to 4 weeks for VITROS Immunodiagnostic products total T4 (TT4) reagent packs. This change affects expired, current (in date) and future product lots.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime