ADVIA Centaur XPT Instrument (2016-06-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
IRL06591516
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA Centaur XPT Instrument10711433

Introduction

Affected products

A. ADVIA Centaur XPT Instrument

Reason

Siemens Healthcare Diagnostics is issuing a customer notification for ADVIA Centaur XPT Waste Reservoir Assembly (SMN 10711091).  Siemens has determined that certain reservoirs manufactured with walls that are below the nominal specification, potentially cause cracking & leaking of seam bottom and leaking at the interface of the seam and reservoir. If leak occurs foaming may occur in the glass jar in the bulk fluid drawer and cause the instrument to stop with low vacuum error #016001324.