Instrument Manager (2016-06-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
driver version 8.00.0017A. Instrument Manager
Manufacturer
Data Innovations LLC
120 Kimball Avenue Suite 100
South Burlington
Vermont
UNITED STATES
VX.XX.XX

Introduction

Affected products

A. Instrument Manager

Reason

Quality control (QC) results without a QC Specimen ID (SID) are being filed with patient results that follow it in the same message on Instrument Manager Siemens DPC Cirrus lMMULITE/ lMMULITE/VersacCell (DPCImmui) driver. If the Cirrus lMMULITE, lMMULITE one, lMMULITE 1000, lMMULITE 2000, lMMULITE 2500, or lMMULITE XPI instrument is running in uni-directional mode and is configured to automatically send QC results to the laboratory information system (LIS) and Specimen IDs are not added to QC samples, then QC results received by IM without a Specimen ID will be added to the first specimen received after these QC results in the same ASTM message.