Health Products

XN-10 MODULE (2016-06-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.  
Manufacturer
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
JAPAN
A. XN-10 MODULEXN-10

Introduction

Affected products

A. XN-10 MODULE

Reason

Sysmex Corporation, Japan (SCJ) determined that some XN-Series analyzers had the wiring for the feeder assembly #113 improperly routed, potentially causing contact with moving parts of the tube sampling mechanism. Contact with the piercer up/down mechanism can cause abrasion of the wire's coating and a short circuit of the tube sampling mechanism.