Health Products

ALUMINUM REACHER ARM (NEW MANITOBA STYLE) (2016-06-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not ApplicableA. ALUMINUM REACHER ARM (NEW MANITOBA STYLE)
Manufacturer
PRISM MEDICAL LTD
485 MILLWAY AVENUE, UNIT 2,
Concord
L4K 3V4
Ontario
CANADA
360404M

Introduction

Affected products

A. ALUMINUM REACHER ARM (NEW MANITOBA STYLE)

Reason

There is a potential for the Reacher arm to "teeter" at the edge of the base of the trolley eyelet that extends from the ceiling track.  When approaching the trolley eyelet with the Reacher arm hook from an angle teetering may occur.  When it occurs the Reacher arm may slide all the way onto the hook (proper final positioning) or slide away from the hook (resulting in falling to the surface below).  This is limited to the customers in the Manitoba and Quebec regions with the newly designed Reacher arm.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime