A. RE-STERILZED ANGIODYNAMICS SOFT-VU OMNI FLUSH ANGIOGRAPHIC CATHETER
Manufacturer
STRYKER SUSTAINABILITY SOLUTIONS
1810 West Drake Dr.
Tempe
85283
Arizona
UNITED STATES
10732203
Introduction
Affected products
A. RE-STERILZED ANGIODYNAMICS SOFT-VU OMNI FLUSH ANGIOGRAPHIC CATHETER
Reason
It was reported to Stryker Sustainability Solutions that the tip of two angiographic catheter's model # 10732203 broke off, prior to being used on the patient. During the complaint investigation, visual inspection confirmed that each device had fractured near the distal tip.