Health Products

RE-STERILZED ANGIODYNAMICS SOFT-VU OMNI FLUSH ANGIOGRAPHIC CATHETER (2016-06-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
253202A. RE-STERILZED ANGIODYNAMICS SOFT-VU OMNI FLUSH ANGIOGRAPHIC CATHETER
Manufacturer
STRYKER SUSTAINABILITY SOLUTIONS
1810 West Drake Dr.
Tempe
85283
Arizona
UNITED STATES
10732203

Introduction

Affected products

A. RE-STERILZED ANGIODYNAMICS SOFT-VU OMNI FLUSH ANGIOGRAPHIC CATHETER

Reason

It was reported to Stryker Sustainability Solutions that the tip of two angiographic catheter's model # 10732203 broke off, prior to being used on the patient. During the complaint investigation, visual inspection confirmed that each device had fractured near the distal tip.