Health Products

POWER PRO AMBULANCE COT (2016-05-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
150840805
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
Michigan
UNITED STATES
B. POWER PRO IT AMBULANCE COT6516000000
More than 10 numbers, contact manufacturer.
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
Michigan
UNITED STATES
C. POWER PRO AMBULANCE COT6506000000
Not applicable
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
Michigan
UNITED STATES
A. CONTROL BOARD FOR POWER PRO AMBULANCE COT6500002014

Introduction

Affected products
  1. CONTROL BOARD FOR POWER PRO AMBULANCE COT
  2. POWER PRO IT AMBULANCE COT
  3. POWER PRO AMBULANCE COT

Reason

A few service reports alleged that "smoke" was coming from the foot end of the cot. This reportedly occurred when either the battery was placed into the cot or shortly after placing the battery into the cot. In addition to the alleged smoking, one of the reports had also noted that when the user attempted to lower the cot (prior to the alleged smoking) the device went into high speed retract and then a few moments later it began to smoke.