POCKET ADAPTOR KIT FOR DEEP BRAIN STIMULATION (2016-06-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
A. POCKET ADAPTOR KIT FOR DEEP BRAIN STIMULATION64001

Introduction

Affected products

A. POCKET ADAPTOR KIT FOR DEEP BRAIN STIMULATION

Reason

Notification communicates recent results from returned product analysis of Medtronic Neuromodulation Deep Brain Stimulation (DBS) system pocket adaptors with reports of high impedances, and reinforces device labeling specific to the handling of pocket adaptors and system integrity checking during implant procedures.

Medtronic has identified sixteen (16) DBS pocket adaptors that were returned for high impedance measurements and where subsequent Medtronic analysis identified conductor wire fractures in close proximity to the location where the adaptor wire exits the neurostimulator connector block.

Notification raises awareness of current product labeling that could reduce potential for pocket adaptor conductor wire fractures or increase detection of an issue during the implant procedure.