Notification communicates recent results from returned product analysis of Medtronic Neuromodulation Deep Brain Stimulation (DBS) system pocket adaptors with reports of high impedances, and reinforces device labeling specific to the handling of pocket adaptors and system integrity checking during implant procedures.
Medtronic has identified sixteen (16) DBS pocket adaptors that were returned for high impedance measurements and where subsequent Medtronic analysis identified conductor wire fractures in close proximity to the location where the adaptor wire exits the neurostimulator connector block.
Notification raises awareness of current product labeling that could reduce potential for pocket adaptor conductor wire fractures or increase detection of an issue during the implant procedure.