Philips DuraDiagnost X-ray System (2016-06-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, General Public, Healthcare Professionals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
SN150002
Manufacturer
PHILIPS HEALTHCARE (SUZHOU) CO., LTD
NO. 258, ZHONG YUAN ROAD
SUZHOU
215024
CHINA
A. Philips DuraDiagnost X-ray System9897-010-02021

Introduction

Affected products

A. Philips DuraDiagnost X-ray System

Reason

DuraDiagnost systems provide an audible exposure termination signal (beep) through the speaker integrated into the monitor. Philips has become aware that buttons or keys on the monitor allow the user to turn the volume setting down to "0", in which case the beep will not be heard at the conclusion of an exposure. Note that the current volume setting is displayed when the buttons or keys are depressed.