B. ALLURA XPER FD20 (PRODUCT OF ALLURACLARITY FAMILY)
722028
102
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
C. ALLURA XPER FD20/20 (PRODUCT OF ALLURACLARITY FAMILY)
722038
271
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
A. ALLURA XPER FD10 (PRODUCT OF ALLURACLARITY FAMILY)
722026
Introduction
Affected products
ALLURA XPER FD10 (PRODUCT OF ALLURACLARITY FAMILY)
ALLURA XPER FD20 (PRODUCT OF ALLURACLARITY FAMILY)
ALLURA XPER FD20/20 (PRODUCT OF ALLURACLARITY FAMILY)
Reason
Philips Healthcare has discovered through trend analysis an increase in the failure rate of certain low-voltage DC power supplies ("DCPS") used in these products. Each system contains multiple DCPS, some of which may be subject to an increased probability of failure. Failure of a DCPS may result in the sudden loss of imaging functionality or mechanical movement, depending on what subsystems the DCPS is powering.