Health Products

PRODUCTS OF ALLURACLARITY FAMILY (2016-06-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
663
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
B. ALLURA XPER FD20 (PRODUCT OF ALLURACLARITY FAMILY)722028
102
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
C. ALLURA XPER FD20/20 (PRODUCT OF ALLURACLARITY FAMILY)722038
271
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
A. ALLURA XPER FD10 (PRODUCT OF ALLURACLARITY FAMILY)722026

Introduction

Affected products
  1. ALLURA XPER FD10 (PRODUCT OF ALLURACLARITY FAMILY)
  2. ALLURA XPER FD20 (PRODUCT OF ALLURACLARITY FAMILY)
  3. ALLURA XPER FD20/20 (PRODUCT OF ALLURACLARITY FAMILY)

Reason

Philips Healthcare has discovered through trend analysis an increase in the failure rate of certain low-voltage DC power supplies ("DCPS") used in these products. Each system contains multiple DCPS, some of which may be subject to an increased probability of failure. Failure of a DCPS may result in the sudden loss of imaging functionality or mechanical movement, depending on what subsystems the DCPS is powering.