Health Products

HEATER COOLER SYSTEM 3T (2016-06-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
Sorin Group Deutschland GMBH
LINBERGHSTRASSE 25
MUNICH
80939
GERMANY
A. HEATER COOLER SYSTEM 3T"16-02-81

Introduction

Affected products

A. HEATER COOLER SYSTEM 3T

Reason

LivaNova has determined that certain oxygenators containing a heat-exchanger with a polymer membrane separating the blood compartment from the water compartment, when used with the cleaning an disinfection regimen described in the IFU for the heater cooler system 3T (IFU versions applicable since 2013), may result in diffusion of a small quantity of hydrogen peroxide from the water side to the blood side during a typical procedure, in a quantity that exceeds the allowable limits determined by LivaNova.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime