Health Products

ADAPTOR AND LIGHT ADAPTOR FOR SMALL BATTERY DRIVE AND SMALL BATTERY DRIVE II (2016-06-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
SYNTHES USA LLC
1302 WRIGHTS LANE EAST
WEST CHESTER
19380
UNITED STATES
A. ADAPTOR AND LIGHT ADAPTOR FOR SMALL BATTERY DRIVE AND SMALL BATTERY DRIVE II"05.001.024

Introduction

Affected products

A. ADAPTOR AND LIGHT ADAPTOR FOR SMALL BATTERY DRIVE AND SMALL BATTERY DRIVE II

Reason

DePuy Synthes has determined there is a potential for the adaptor and light adaptor for Small Battery Drive and Small Battery Drive II to generate excessive internal pressure that may cause the respective products to burst. Although preliminary internal tests for this failure mode have determined this scenario is highly unlikely to occur, DePuy Synthes has decided to recall the products.