Health Products

Smith & Nephew, Inc. Hip Replacement Parts (Switzerland) (2016-06-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.H. POLARCUP CORTICAL SCREW - S&N
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75017232
More than 10 numbers, contact manufacturer.I. SLR-PLUS STANDARD FEMORAL STEMS (S+N)
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75002795
More than 10 numbers, contact manufacturer.O. POLARCUP PE X-LINKED INSERT
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75018942
More than 10 numbers, contact manufacturer.F. POLARCUP PE INSERT - S&N
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
75017207
More than 10 numbers, contact manufacturer.G. POLARCUP ANCHORING PEG - S&N
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
75017231
More than 10 numbers, contact manufacturer.P. SL-PLUS INTEGRATION-PLUS MIA STEM LATERAL WITH TI/HA
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75000188
More than 10 numbers, contact manufacturer.D. POLARCUP SHELL HA - ORTHO
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"11000317
More than 10 numbers, contact manufacturer.E. POLARCUP SHELL CEMENTED - S&N
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75017194
More than 10 numbers, contact manufacturer.K. POLARSTEM STANDARD TI-PLASMA/HA SAP #
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"71500462
More than 10 numbers, contact manufacturer.B. 529750  SL-PLUS LATERAL CEMENTLESS FEMORAL HIP STEM
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75002748
More than 10 numbers, contact manufacturer.C. POLARCUP SHELL TI-PLASMA - S&N
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75017183
More than 10 numbers, contact manufacturer.L. POLARSTEM LATERAL TI-PLASMA/HA  SAP #
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"71500474
More than 10 numbers, contact manufacturer.M. SL-PLUS MIA STANDARD
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75001938
More than 10 numbers, contact manufacturer.N. ECHELON HIP SYSTEM-TITANIUM PRIMARY STD OFFSET
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75001952
More than 10 numbers, contact manufacturer.A. SL-PLUS STANDARD CEMENTLESS FEMORAL HIP STEM
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75002695
More than 10 numbers, contact manufacturer.J. SLR-PLUS LATERAL FEMORAL STEMS (S+N)
Manufacturer
SMITH & NEPHEW ORTHOPAEDICS AG
OBERNEUHOFSTRASSE 10D
BAAR
6340
SWITZERLAND
"75001927

Introduction

Affected products
  1. SL-PLUS STANDARD CEMENTLESS FEMORAL HIP STEM
  2. 529750 SL-PLUS LATERAL CEMENTLESS FEMORAL HIP STEM
  3. POLARCUP SHELL TI-PLASMA - S&N
  4. POLARCUP SHELL HA - ORTHO
  5. POLARCUP SHELL CEMENTED - S&N
  6. POLARCUP PE INSERT - S&N
  7. POLARCUP ANCHORING PEG - S&N
  8. POLARCUP CORTICAL SCREW - S&N
  9. SLR-PLUS STANDARD FEMORAL STEMS (S+N)
  10. SLR-PLUS LATERAL FEMORAL STEMS (S+N)
  11. POLARSTEM STANDARD TI-PLASMA/HA SAP #
  12. POLARSTEM LATERAL TI-PLASMA/HA  SAP #
  13. SL-PLUS MIA STANDARD
  14. ECHELON HIP SYSTEM-TITANIUM PRIMARY STD OFFSET
  15. POLARCUP PE X-LINKED INSERT
  16. SL-PLUS INTEGRATION-PLUS MIA STEM LATERAL WITH TI/HA

Reason

Smith & Nephew, Inc. has initiated a voluntary field safety corrective action/recall of the ceramic hip prostheses. The current hip prostheses IFUs do not provide adequate instruction on the revision surgeries due to ceramic failure. The IFU does not provide the detailed instructions that if a hip prostheses revision is required due to a fractured ceramic component, revisions shall only be performed using a ceramic femoral ball head in order to prevent any potential undue wear of the revision components.