Health Products

BD Trucount tubes (2016-06-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
6083708
Manufacturer
Becton, Dickinson and Company
BD Biosciences
2350 Qume Drive
San Jose
95131
California
UNITED STATES
A. BD Trucount tubes340334

Introduction

Affected products

A. BD Trucount tubes

Reason

Pouches contained in the kit lot 6083708 may not be sealed properly. The pouches contained in the kit can be identified with the lot no.16040 and are embossed with "17 02 16 t3" or "16 02 16 T3". Improper storage of Trucount Tubes due to failed pouch closure can affect the performance of the counting beads leading to erroneous results in enumeration assays utilizing Trucount. If the pouch seal is intact and the desiccant included in the pouch is blue, the product may be used. If the pouch seal integrity is compromised and/or the desiccant has turned from blue to lavender, discard the tubes, because exposure to the ambient environment can rapidly degrade the performance of the product.