Health Products

cobas Urinalysis System and Microscopy Analyzer (2016-07-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Software v. less or eq 2.2.1.
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
B. COBAS 6500 - COBAS U701 MICROSCOPY ANALYZER6390501001
Software v. less or eq 2.2.1.
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
A. COBAS 6500 U601 URINALYSIS SYSTEM - URINE ANALYZE6390498001

Introduction

Affected products
  1. COBAS 6500 U601 URINALYSIS SYSTEM - URINE ANALYZE
  2. COBAS 6500 - COBAS U701 MICROSCOPY ANALYZER

Reason

During internal investigation, our manufacturer confirmed a possible software issue where the strips cassettes (cobas u pack) on cobas u 601 urinalysis system is denied by the user software although the expiry date has not been reached. This is true for every month reaching the cassette expiry date.