Health Products

IOL Master 700 (2015-02-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1128467
Manufacturer
Carl Zeiss Meditech AG
Goeschwitzer Strasse 51-52
Jena
07745
GERMANY
A. IOL Master 700IOL MASTER 700

Introduction

Affected products

A. IOL Master 700

Reason

The current IOLmaster 700 software version 1.14.84 sometimes shows an unexpected behavior during toric IOL power calculation with the haigis suite: in some cases of low cylinder power with a theoretically ideal IOL below the delivery range of the selected toric IOL model, the device suggests a toric IOL with significant remaining cylinder refraction.