B. AllSet Gold SSP HLA Class I High Resolution Kits
Manufacturer
One Lambda Inc
21001 Kittridge Street
Canoga Park
California
UNITED STATES
541100D
Lot 22, Batch 1256853
C. AllSet Gold HLA Low Resolution Kits
Manufacturer
One Lambda Inc
21001 Kittridge Street
Canoga Park
California
UNITED STATES
54310D
Lot 12, Batch 1303557
A. AllSet Gold SSP HLA Direct to High Resolution Kit
Manufacturer
One Lambda Inc
21001 Kittridge Street
Canoga Park
California
UNITED STATES
54010D
Introduction
AllSet Gold SSP HLA Direct to High Resolution Kit
AllSet Gold SSP HLA Class I High Resolution Kits
AllSet Gold HLA Low Resolution Kits
AllSet + Gold SSP
AllSet Gold ABDR Low Resolution Kit
AllSet Gold ABDRQ Low Resolution Kit
Reason
A false negative reaction may be observed in the presence of a*11:01:04, specifically for primers PM118C, PM118F, A11-01 and A11-01A. This could lead to mis-typings during analysis