Health Products

MOSAIQ (2016-07-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
VERSION 2.50.05 AND HIGHER
Manufacturer
Elekta Business Area Software Systems IMPAC Medical Systems, Inc.
A. MOSAIQMOSAIQ

Introduction

Affected products

A. MOSAIQ

Reason

Retention of outdated information in an order set. It is possible that a change to an order set will not be saved in the current open care plan when the following work flow is utilized:
  1. open either the care plan edit form or the care plan instance form and leave the form open.
  2. select an order set that is part of the same care plan and make a change to the order set. This can be changed to the order set name and label or a change to the drug, dose, administration instructions, etc.
  3. the update made to the order set will not be saved in the current open care plan.