Health Products

MULTIDIAGNOST ELEVA IMAGE DETECTORS (2016-07-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
B. MULTIDIAGNOST ELEVA - FLAT DETECTOR - IMAGE DETECTOR0708MD02
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
A. MULTIDIAGNOST ELEVA - IMAGE INTENSIFIER - TV IMAGE DETECTOR0708MD02

Introduction

Affected products
  1. MULTIDIAGNOST ELEVA - IMAGE INTENSIFIER - TV IMAGE DETECTOR
  2. MULTIDIAGNOST ELEVA - FLAT DETECTOR - IMAGE DETECTOR

Reason

Philips has become aware that the two bolts that attach the clamping plate to the lateral shaft may break. This may cause the c-arm to float freely. Depending on the orientation of the free-floating C-arm and any other forces applied to it the C-arm could move laterally forward, slide out of its bearings and then possibly sag.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime