2015006968A. ENA HRI TIPS
Manufacturer
MICERIUM S.P.A
ViA G MARCONI 83
AVEGNO, GENOVA
16036
ITALY
EHRTUD2

Introduction

Affected products

A. ENA HRI TIPS

Reason

Material error of the tips has been determined as the cause. The injection mold for producing the tips is damaged by wear and tear at certain points, so that the plastic shows reduced mechanical stability.