A specimen misidentification may occur, despite an active software prompt on the screen, when the following sequence of events occurs:
pending test orders exist on the worklist
a new specimen ID is added using the "next" icon on the sample analysis - patient results screen or by scanning a specimen barcode label
the user leaves the screen before aspirating the new specimen
upon returning to the sample analysis - patient results screen, the following prompt displays "worklist contains entries. Confirm next specimen ID is correct before processing specimen"
the next specimen ID field of the DxH 500 defaults to the first specimen ID on the worklist which may not be the newest specimen identified
the issue occurs when the user fails to verify the specimen ID
The issue does not occur when the worklist has no pending test orders.