POLESTAR N-20 - MAGNETIC RESONANCE DIAGNOSTIC DEVICE (2016-06-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
3044
Manufacturer
Medtronic Navigation Israel Ltd.
3 HACARMEL STREET
KOCHAV YOKNEAM BUILDING
ISRAEL
A. POLESTAR N-20 - MAGNETIC RESONANCE DIAGNOSTIC DEVICEPOLESTAR N-20

Introduction

Affected products

A. POLESTAR N-20 - MAGNETIC RESONANCE DIAGNOSTIC DEVICE

Reason

Notification related to a potential risk for warning lamp connectors and/or chiller grounding discontinuity. Manufacturer identified during in-house testing a potential ground discontinuity in sub-assemblies, leading to potential risk of user electrical shock upon direct contact with these sub-assemblies. Risk of user electrical shock was low. In addition the system residual risk levels did not change and remained acceptable. Manufacturer also determined that there was no change in risk to the patient since the assemblies were located outside the OR.