Health Products

XPRECIA STRIDE COAGULATION ANALYZER (2016-07-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
100889
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. XPRECIA STRIDE COAGULATION ANALYZER10714596

Introduction

Affected products

A. XPRECIA STRIDE COAGULATION ANALYZER

Reason

Siemens has confirmed that version (v1.0) of the data management software (DMS) CD was packaged with some Xprecia Stride Coagulation Analyzers. The version of the DMS CD should have been v1.1, as this will match the version of firmware pre-loaded on the analyzer. These two versions must match in order to provide full functionality to the analyzer.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime