Health Products

ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - COLI (2016-07-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.  
Manufacturer
Biomerieux Sa
Chemin De L'orme
Marcy-l'etoile
69280
FRANCE
A. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - COLI537300

Introduction

Affected products
  1. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - COLI

Reason

Based on scientific publications (including EUCAST warning letter on their website) on colistin gradient and disk diffusion tests, including Etest colistin, bioMerieux was alerted about a potential underestimation of the MIC values for entero bacteriaceae and acinetobacter. bioMerieux initiated a complaint investigation to confirm product performance and determine root cause. While the internal and external investigations are still ongoing.