Health Products

INTELLIVUE PATIENT MONITOR (2016-07-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Hospitals, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not Applicable
Manufacturer
Philips Medizin Systeme Boeblingen GMBH
HEWLETT-PACKARD STR. 2,
BOEBLINGEN
71034
GERMANY
D. INTELLIVUE MX550 PATIENT MONITOR866066
Not Applicable
Manufacturer
Philips Medizin Systeme Boeblingen GMBH
HEWLETT-PACKARD STR. 2,
BOEBLINGEN
71034
GERMANY
B. INTELLIVUE MX500 PATIENT MONITOR866064
Not Applicable
Manufacturer
Philips Medizin Systeme Boeblingen GMBH
HEWLETT-PACKARD STR. 2,
BOEBLINGEN
71034
GERMANY
C.  INTELLIVUE MX450 PATIENT MONITOR866062
Not Applicable
Manufacturer
Philips Medizin Systeme Boeblingen GMBH
HEWLETT-PACKARD STR. 2,
BOEBLINGEN
71034
GERMANY
A. INTELLIVUE MX400 PATIENT MONITOR866060

Introduction

 Affected products
  1. INTELLIVUE MX400 PATIENT MONITOR
  2. INTELLIVUE MX500 PATIENT MONITOR
  3. INTELLIVUE MX450 PATIENT MONITOR
  4. INTELLIVUE MX550 PATIENT MONITOR

Reason

Philips has recently determined that when certain patient monitors (mx400, mx450, mx500 and mx550) have been powered on continuously for at least three months, the waveforms displayed on the monitor may be outdated and therefore fail to reflect the patient's current condition. Should this occur, however, the monitor's alarms and numeric displays of vital signs will continue to function as specified and accurately reflect current patient data.