Contamination of ARTIS Systems (2016-07-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Healthcare Professionals, General Public, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
A. 53081
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
A. AXIOM ARTIS DBAA. 7555357
A. 53081
Manufacturer
CONTINENTAL SURGICAL
111 ALBERT ST
OTTAWA
K1P 1B1
Ontario
CANADA
A. AXIOM ARTIS DBAA. 7555357

Introduction

Affected products
  1. AXIOM ARTIS DBA
  2. AXIOM ARTIS DTA
  3. AXIOM ARTIS DBC MAGNETIC NAVIGATION
  4. ARTIS ZEE FLOOR MN
  5. PROXIMAL HUMERAL PLATE EXTRA LONG
  6. ARTIS ZEE MULTI-PURPOSE
  7. ARTIS ZEE FLOOR
  8. ARTIS ZEE CEILING
  9. ARTIS ZEE BIPLANE
  10. AXIOM ARTIS FC
  11. AXIOM ARTIS DFC MAGNETIC NAVIGATION
  12. AXIOM ARTIS DFC
  13. AXIOM ARTIS FA ALPHA
  14. AXIOM ARTIS MP DIAGNOSTIC
  15. AXIOM ARTIS BC
  16. AXIOM ARTIS BA
  17. AXIOM ARTIS TA
  18. ARTIS Q
  19. ARTIS Q.ZEN

Reason

Siemens is initiating a field corrective action, as during routine surveillance activities, it was identified that contaminants in the form of biomass can develop in the cooling system of Artis systems, which can result in damage to the pump system. Biomass is developing in the ongoing life time of a system.