Contamination of ARTIS Systems (2016-07-27) Jul 27, 2016 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : Healthcare Professionals, General Public, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number A. 53081 Manufacturer Siemens Healthcare GMBH HENKESTRASSE 127 ERLANGEN 91052 GERMANY A. AXIOM ARTIS DBA A. 7555357 A. 53081 Manufacturer CONTINENTAL SURGICAL 111 ALBERT ST OTTAWA K1P 1B1 Ontario CANADA A. AXIOM ARTIS DBA A. 7555357
Affected products
AXIOM ARTIS DBA
AXIOM ARTIS DTA
AXIOM ARTIS DBC MAGNETIC NAVIGATION
ARTIS ZEE FLOOR MN
PROXIMAL HUMERAL PLATE EXTRA LONG
ARTIS ZEE MULTI-PURPOSE
ARTIS ZEE FLOOR
ARTIS ZEE CEILING
ARTIS ZEE BIPLANE
AXIOM ARTIS FC
AXIOM ARTIS DFC MAGNETIC NAVIGATION
AXIOM ARTIS DFC
AXIOM ARTIS FA ALPHA
AXIOM ARTIS MP DIAGNOSTIC
AXIOM ARTIS BC
AXIOM ARTIS BA
AXIOM ARTIS TA
ARTIS Q
ARTIS Q.ZEN
Siemens is initiating a field corrective action, as during routine surveillance activities, it was identified that contaminants in the form of biomass can develop in the cooling system of Artis systems, which can result in damage to the pump system. Biomass is developing in the ongoing life time of a system.