Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
24601A. YSIO MAX
Manufacturer
Siemens Healthcare GMBH
HENKESTRASSE 127
ERLANGEN
91052
GERMANY
10762470

Introduction

Affected products
  1. YSIO MAX

Reason

Siemens identified an issue with a patient table which was caused by a manufacturing error. There is an unlikely risk that an internal part of the table may detach causing the table top to tilt to the left side which could result in an injury to a patient on the table. This issue was identified during the installation of a patient table and this is the only known table with this problem. Siemens is initiating steps to determine if the same problem exists on other patient tables.