Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
A. UNICEL DXH SLIDEMAKER STAINER775222

Introduction

Affected products

A. UNICEL DXH SLIDEMAKER STAINER

Reason

Beckman Coulter has received and confirmed a report of a fire within the Stainer module of a customer's DxH Slidemaker Stainer. The fire was fully contained within the stainer module. The event was preceded by a significant increase in the frequency of the automatic top-off of the module's methanol bath. Beckman Coulter is initiating this field action to reduce the potential for flammable liquids coming into contact with electrical components in the stainer module possibly causing a fire.