UNICEL DXH 800/600 COULTER CELLULAR ANALYSIS SYSTEM (2016-08-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: Hospitals, General Public, Healthcare Professionals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
B. UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEMB23858
All lots
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
A. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM629029

Introduction

Affected products
  1. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
  2. UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEM

Reason

Results from a patient control sample could be misidentified by being associated with a pending test order for a patient sample already on the worklist when the following sequence of events occurs:
- the primary identifier is tube position ID
- a patient control file is setup for ID xxxxx, and a pending test order for a patient sample ID xxxxx is subsequently added to the worklist
- the patient control sample is analyzed using cassette presentation, but that tube does not contain a barcode-labeled specimen ID or the label cannot be read.