B. UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEM
B23858
All lots
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
A. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
629029
Introduction
Affected products
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEM
Reason
Results from a patient control sample could be misidentified by being associated with a pending test order for a patient sample already on the worklist when the following sequence of events occurs:
- the primary identifier is tube position ID
- a patient control file is setup for ID xxxxx, and a pending test order for a patient sample ID xxxxx is subsequently added to the worklist
- the patient control sample is analyzed using cassette presentation, but that tube does not contain a barcode-labeled specimen ID or the label cannot be read.