Siemens Healthcare Diagnostics has confirmed that the ADVIA Centaur systems myoglobin assay is not meeting the analytical sensitivity claim of < 3 ng/ml (ug/l) as specified in the ADVIA Centaur myoglobin instructions for use (IFU), for all in-date reagent lots when evaluated with calibrator U kit lots ending in 63 and 64. Values up to 12 ng/ml (ug/l) were observed.
Siemens' investigation has identified that the calibrator U kit lots ending in 63 or 64 have drifted from the internal standardization causing a positive shift in results.