Health Products

SPIRIT PLUS AND SPIRIT SELECT BED (2016-07-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.  
Manufacturer
CHG HOSPITAL BEDS INC
1020 ADELAIDE STREET SOUTH
LONDON
N6E 1R6
Ontario
CANADA
SPIRIT PLUS AND SPIRIT SELECT BED"N2F02BEMP1

Introduction

Affected products
  1. SPIRIT PLUS AND SPIRIT SELECT BED

Reason

The spirit plus and spirit select beds have actuators, called the hi-lo actuators, that allow for the bed to be raised and lowered. Each bed has two actuators, one that controls the headend of the bed's movement and one for the foot end. The actuators are mounted to the frame by use of a clevis pin design. The beds lower to a minimum height of 10.75 inches and a maximum height of 34 inches. It was identified that in some situations the actuator may break at the mount end.