Health Products

Infant Warmer system (2016-08-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Ohmeda Medical, A Division Of Datex Ohmeda Inc.
B. Infant Warmer System (IWS)2001
Not Applicable
Manufacturer
Ohmeda Medical, A Division Of Datex Ohmeda Inc.
Infant Warmer system3051

Introduction

Affected products
  1. Infant Warmer system
  2. Infant Warmer System (IWS)

Reason

Hot screws from the heater head of the infant warmer system could fall onto the bed if the "heater head" assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime