The Medline packs contain bard irrigation syringes under recall Bard. Bard is recalling the syringes because the syringe may be at risk for a slit defect in the packaging that may affect sterility. The Bard syringe in these packs is to be removed and discarded. The Medline packs are terminally sterilized, therefore the remaining components within the packs are safe to use.