Hospira MedNet Medication Management Suite - with SapphirePlus 13.1x Infusion Pump (2016-07-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Version 6.1
Manufacturer
Hospira Inc.
275 North Field Drive
Lake Forest
60045
Illinois
UNITED STATES
Hospira MedNet Medication Management Suite - with SapphirePlus 13.1x Infusion Pump16037-64-03

Introduction

Affected Products

Hospira MedNet Medication Management Suite - with SapphirePlus 13.1x Infusion Pump

Reason

Update from the manufacturer on January 11th, 2017 :  Hospira Healthcare Corporation issued an update to the letter issued on July 28th 2016. Hospira is issuing this update to notify users incorrect calculations can also occur under the following conditions:

- Medication concentration is listed in Micrograms (mcg), bolus is enabled and a Maximum Bolus Amount listed in nanograms (nanog) or    
- Medication concentration is listed in Units, bolus is enabled and Maximum Bolus Amount is listed in milliUnits.

In these situations, the infuser incorrectly calculates the maximum bolus amount. This combination of clinical parameters within the drug library allows for the potential of an incorrect programming for the bolus amount that could be greater than what was intended for the maximum bolus amount. Medications whose drug library entries do not involve Maximum Bolus Amount units in Nanogram (nanog) or milliUnits are not affected by this issue.

There is a potential for software issue in MedNet version 6.1 when used with Sapphire Plus 13.1x Infusion Pumps. When a medication using a dose rate unit of million units, and bolus is enabled in the drug library for that medication, and concentration is not equal to 1 million units / ml, then the bolus amount calculated on the Sapphire Plus Pump will be incorrect.