Health Products

TANGO AUTOMATED BLOOD BANK ANALYZER SYSTEM - Tango Infinity (2016-07-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Bio-Rad Medical Diagnostics GMBH
Industriestr.1
Dreieich
TANGO AUTOMATED BLOOD BANK ANALYZER SYSTEM - Tango Infinity850000010

Introduction

Affected Products

TANGO AUTOMATED BLOOD BANK ANALYZER SYSTEM - Tango Infinity

Reason

A microplate was inadvertantly placed in the loading station while in function, displacing the previous plate. A erroneous test result was attributed to the placement of the wrong microplate.