MULTIDIAGNOST ELEVA, OMINIDOIAGNOST ELEVA, ALLURA XPER VARIOUS MODELS

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
H. ALLURA XPER FD20 SYSTEM - MAIN UNITALLURA XPER FD20
146,33
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
I. ALLURA XPER FD20 OR TABLE454
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
F. OMNIDIAGNOST ELEVA708027
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
G. ALLURA XPER FD10ALLURA XPER FD10
168
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
D. ALLURA XPER FD20/10 SYSTEM - MAIN UNITALLURA XPER FD20/10
171
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
E. ALLURA XPER FD20/20ALLURA XPER FD20/20
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
B. MULTIDIAGNOST ELEVA - FLAT DETECTOR - IMAGE DETECTOR0708MD02
154
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
C. ALLURA XPER FD10/10 - MAIN UNITALLURA XPER FD10/10
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
A. MULTIDIAGNOST ELEVA - IMAGE INTENSIFIER - TV IMAGE DETECTOR0708MD02

Introduction

Affected Products
  1. MULTIDIAGNOST ELEVA - IMAGE INTENSIFIER - TV IMAGE DETECTOR
  2. MULTIDIAGNOST ELEVA - FLAT DETECTOR - IMAGE DETECTOR
  3. ALLURA XPER FD10/10 - MAIN UNIT
  4. ALLURA XPER FD20/10 SYSTEM - MAIN UNIT
  5. ALLURA XPER FD20/20
  6. OMNIDIAGNOST ELEVA
  7. ALLURA XPER FD10
  8. ALLURA XPER FD20 SYSTEM - MAIN UNIT
  9. ALLURA XPER FD20 OR TABLE

Reason

Philips Healthcare has discovered through customer feedback, a system restart failure as a result of a blown fuse. The fuse was blown due to an unexpected main power failure in the hospital in which the uninterruptible power supply (UPS) took over.

Philips discovered after internal investigation that in time the fuse degrades.

The degrading depends on the initial value of the fuse, the current through the fuse, a continuous high operating temperature, the environmental temperature and the time the system is operational.

After degradation, the blowing of the fuse may be triggered by an unexpected main power failure as described above.