Health Products

DIMENSION VISTA AND EMIT II OPIATE ASSAYS

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
EA6091
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark, Delaware
19714-6101
UNITED STATES
B. DIMENSION CLINICAL CHEMISTRY SYSTEM - OPIATES (UOPI)DF93A
H2
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark, Delaware
19714-6101
UNITED STATES
C. EMIT II PLUS OPIATES ASSAY9B309UL
15078BB
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark, Delaware
19714-6101
UNITED STATES
A. DIMENSION VISTA SYSTEM - URINE OPIATES SCREENK5093

Introduction

Affected Products
  1. DIMENSION VISTA SYSTEM - URINE OPIATES SCREEN
  2. DIMENSION CLINICAL CHEMISTRY SYSTEM - OPIATES (UOPI)
  3. EMIT II PLUS OPIATES ASSAY

Reason

Siemens investigation confirmed that the introduction of a new SYVA polyclonal antibody lot (G1) for use in the dimension urine opiates screen has resulted in an alteration of specificity for Levallorphan, Levorphanol, Nalorphine and Oxycodone at the 300 and 2000 NG/ml cutoff concentration. As a result in the change in sensitivity to the cross reactant, the claims for these compounds are no longer accurate for either the 300 or 2000 NG/ml cutoff as stated within the instructions for use (IFU). Further investigation confirmed that SYVA, Beckman Coulter and Advia Chemistry Opiate reagents are impacted as well due to a common raw material used in the assays.