The following issue is being addressed by GE Healthcare file no. 25470. The Revolution CT exceeds IEC conducted emission requirements. During recent certification testing, GE Healthcare determined that one configuration of the Revolution CT product exceeds the conducted emissions limits specified in the IEC60601-1-2:2007 EMC standard. Medical devices powered on the same power line as the Revolution CT Power Distribution Unit (PDU) may be affected by electrical emissions through the power line. Equipment powered from the CT System such as the EKG Monitor or Patient Contrast Injector is not affected.