Revolution CT- Main Unit (2016-08-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
5445060CN7
Manufacturer
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
Revolution CT- Main Unit55590000

Introduction

Affected Products

Revolution CT- Main Unit

Reason

The following issue is being addressed by GE Healthcare file no. 25470. The Revolution CT exceeds IEC conducted emission requirements. During recent certification testing, GE Healthcare determined that one configuration of the Revolution CT product exceeds the conducted emissions limits specified in the IEC60601-1-2:2007 EMC standard. Medical devices powered on the same power line as the Revolution CT Power Distribution Unit (PDU) may be affected by electrical emissions through the power line. Equipment powered from the CT System such as the EKG Monitor or Patient Contrast Injector is not affected.