Health Products

Brilliance 40/64 Slice CT System, Ingenuity CT System (2016-08-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
30003B. Ingenuity CT System
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES
728323
10213A. Brilliance 40/64 Slice CT System
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES
4550 110 09021

Introduction

Affected Products
  1. Brilliance 40/64 Slice CT System
  2. Ingenuity CT System

Reason

Philips has become aware that while performing high resolution head scans, the reconstructed images may exhibit degraded image quality manifested as:
  • non uniformity of the image
  • CT number shift, greater than 5 Hounsfield units
  • reduction in gray/white matter differentiation


the issue is most severe in the high resolution mode. Philips provides factory reference pediatric scan protocols which are set to high resolution by default. Adult head scans have a factory reference protocol that defaults to standard resolution which does not experience the issue