B. Tecnis 1-Piece IOLWith Tecnis LTEC Preloaded Delivery System
PCB00
8496921508
Manufacturer
Abbott Medical Optics, Inc.
1700 E. St. Andrew Place
C. Tecnis Symfony Extended Range Of Vision IOL
ZXR00
2015871503
Manufacturer
Abbott Medical Optics, Inc.
1700 E. St. Andrew Place
A. Sensar Foldable IOL With OptiEdge
AR40E
Introduction
Affected Products
Sensar Foldable IOL With OptiEdge
Tecnis 1-Piece IOLWith Tecnis LTEC Preloaded Delivery System
Tecnis Symfony Extended Range Of Vision IOL
Tecnis 1-Piece Acrylic Aspheric IOL
Reason
AMO is initiating this action due to detection of an inspection equipment malfunction that resulted in a total of 58 distributed lenses not being fully checked prior to release. It has been identified that this malfunction may result in the release of mislabelled intraocular lenses. Use of a mislabelled intraocular lens could lead to potential unexpected postoperative refractive error and may result in a secondary surgical intervention. Intraocular lenses with serial numbers included in the recall letter that have been implanted and resulted in an appropriate post-op refractive outcome are not affected by this action.