Abl90 Flex - Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not Applicable.
Manufacturer
Radiometer Medical Aps
Akandevej 21
Bronshoj
2700
DENMARK
Abl90 Flex - Analyzer393-090

Introduction

Affected Products

Abl90 Flex - Analyzer

Reason

It has been discovered that the feature "exit conditioning" may cause a risk to patients. When a new sensor cassette is installed, the analyzer will automatically run extra calibrations and internal QC's for the first 24 hours. It is possible, during sensor cassette start-up to press "exit conditioning". This will cancel the second start-up calibration and the 3 start-up QC's. It will not affect the remaining frequent calibrations. The analyzer does not indicate the exit from the conditioning and the patient results are not flagged for QC not being run.