It has been discovered that the feature "exit conditioning" may cause a risk to patients. When a new sensor cassette is installed, the analyzer will automatically run extra calibrations and internal QC's for the first 24 hours. It is possible, during sensor cassette start-up to press "exit conditioning". This will cancel the second start-up calibration and the 3 start-up QC's. It will not affect the remaining frequent calibrations. The analyzer does not indicate the exit from the conditioning and the patient results are not flagged for QC not being run.