Health Products

Aquios Cl Flow Cytometer

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
AY03004
Manufacturer
Beckman Coulter Ireland, Inc.
Lismeehan, O'Callaghan's Mills
County Clare
IRELAND
Aquios Cl Flow CytometerB30166

Introduction

Affected Products

Aquios Cl Flow Cytometer

Reason

The use of positions 9 and 10 of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe. Damage to the prep probe will render the instrument inoperative with the potential to delay reporting of results.